Enhanced Response Calibration
Enhanced Response calibration is a simplified calibration method when you only need S11 or S21 measurements. It is faster than full SOLT calibration but provides reduced accuracy.
What You Will Learn
Section titled “What You Will Learn”- When to use Enhanced Response calibration
- S11-only calibration (OPEN or SHORT)
- S21-only calibration (THRU)
- Limitations of simplified calibration
When to Use Enhanced Response
Section titled “When to Use Enhanced Response”Enhanced Response calibration is appropriate when:
- You only need S11 OR S21 (not both simultaneously)
- You want a quick calibration for field work
- You do not have a complete calibration kit
- Absolute accuracy is less critical than relative measurements
S11 Enhanced Response (Reflection Only)
Section titled “S11 Enhanced Response (Reflection Only)”For S11 measurements, you can calibrate with just OPEN or SHORT (not both).
Using OPEN Only
Section titled “Using OPEN Only”-
Set your frequency range
Go to STIMULUS and configure START and STOP frequencies.
-
Open calibration menu
Go to CAL > CALIBRATE
-
Perform OPEN calibration
- Remove the connector from Port 1 or connect an OPEN standard
- Touch OPEN
- Wait for sweep to complete
-
Complete calibration
Touch DONE
Using SHORT Only
Section titled “Using SHORT Only”-
Set your frequency range
-
Open calibration menu
Go to CAL > CALIBRATE
-
Perform SHORT calibration
- Connect SHORT standard to Port 1
- Touch SHORT
- Wait for completion
-
Complete calibration
Touch DONE
S21 Enhanced Response (Transmission Only)
Section titled “S21 Enhanced Response (Transmission Only)”For S21 measurements, you can calibrate with just THRU.
-
Set your frequency range
Go to STIMULUS and configure START and STOP frequencies.
-
Connect cables in THRU configuration
Connect your Port 1 cable directly to your Port 2 cable (or use a THRU standard).
-
Open calibration menu
Go to CAL > CALIBRATE
-
Perform THRU calibration
- Touch THRU
- Wait for sweep to complete
-
Complete calibration
Touch DONE
Enabling Enhanced Response Mode
Section titled “Enabling Enhanced Response Mode”The NanoVNA automatically detects Enhanced Response mode based on which calibration standards were performed.
You can also explicitly enable it:
-
Go to calibration menu
Touch CAL
-
Select Enhanced Response
Touch ENHANCED RESPONSE
This toggles Enhanced Response mode on or off.
-
Check the status indicator
When Enhanced Response is active, the calibration indicator may show differently than full calibration.
Understanding the Limitations
Section titled “Understanding the Limitations”S11 Enhanced Response Limitations
Section titled “S11 Enhanced Response Limitations”| Missing Standard | Impact |
|---|---|
| No LOAD | Directivity error not corrected |
| No SHORT (when using OPEN) | Limited accuracy for low-impedance loads |
| No OPEN (when using SHORT) | Limited accuracy for high-impedance loads |
Practical impact:
- Works well for measuring resonant antennas near 50 ohms
- Less accurate for very low or very high impedances
- Phase accuracy is reduced
S21 Enhanced Response Limitations
Section titled “S21 Enhanced Response Limitations”| Missing Standard | Impact |
|---|---|
| No ISOLN | Port crosstalk not corrected |
| No Port 1 calibration | Source match errors not corrected |
Practical impact:
- Good for measuring insertion loss of attenuators and filters
- Less accurate for devices with poor return loss
- Isolation measurements may show floor at -40 to -50 dB
Quick Reference: Minimum Calibration Requirements
Section titled “Quick Reference: Minimum Calibration Requirements”Minimum: OPEN only or SHORT only
Better: OPEN + SHORT
Best: OPEN + SHORT + LOAD (full S11 calibration)
Use for: Antenna SWR, return loss measurements
Minimum: THRU only
Better: THRU + ISOLN
Best: Full SOLT (OPEN + SHORT + LOAD + THRU)
Use for: Filter insertion loss, cable loss measurements
Minimum: Not recommended
Required: Full SOLT calibration
Use for: Complete device characterization, accurate impedance and transmission measurements
Example: Quick Antenna Check in the Field
Section titled “Example: Quick Antenna Check in the Field”When you need a fast antenna check without a full calibration kit:
-
Set frequency range for your band
Example: 144 MHz to 148 MHz for 2m band
-
Calibrate with OPEN
Remove the antenna connector and perform OPEN calibration.
-
Connect your antenna
-
Read SWR
The SWR reading will be approximate but useful for tuning.
-
For better accuracy
If you have a SHORT standard (even a piece of wire soldered across an SMA connector), perform SHORT calibration as well.
Comparing Calibration Methods
Section titled “Comparing Calibration Methods”| Method | Standards Required | S11 Accuracy | S21 Accuracy | Speed |
|---|---|---|---|---|
| No calibration | None | Poor | Poor | Instant |
| OPEN only | 1 | Fair | N/A | Fast |
| SHORT only | 1 | Fair | N/A | Fast |
| THRU only | 1 | N/A | Fair | Fast |
| OPEN + SHORT | 2 | Good | N/A | Fast |
| THRU + ISOLN | 2 | N/A | Good | Fast |
| Full SOLT | 4-5 | Excellent | Excellent | Moderate |
Switching Between Calibration Modes
Section titled “Switching Between Calibration Modes”-
Check current calibration status
The status bar shows which calibration is active.
-
Recall a different calibration
Go to CAL > menu to access saved calibrations.
-
Reset and recalibrate
Go to CAL > RESET to clear current calibration, then perform new calibration.
Next Steps
Section titled “Next Steps”- Full Calibration - Complete SOLT procedure
- Saving Calibration - Store calibration for reuse
- Your First S11 Measurement - Apply your calibration