Skip to content

Enhanced Response Calibration

Enhanced Response calibration is a simplified calibration method when you only need S11 or S21 measurements. It is faster than full SOLT calibration but provides reduced accuracy.

  • When to use Enhanced Response calibration
  • S11-only calibration (OPEN or SHORT)
  • S21-only calibration (THRU)
  • Limitations of simplified calibration

Enhanced Response calibration is appropriate when:

  • You only need S11 OR S21 (not both simultaneously)
  • You want a quick calibration for field work
  • You do not have a complete calibration kit
  • Absolute accuracy is less critical than relative measurements

For S11 measurements, you can calibrate with just OPEN or SHORT (not both).

  1. Set your frequency range

    Go to STIMULUS and configure START and STOP frequencies.

  2. Open calibration menu

    Go to CAL > CALIBRATE

  3. Perform OPEN calibration

    • Remove the connector from Port 1 or connect an OPEN standard
    • Touch OPEN
    • Wait for sweep to complete
  4. Complete calibration

    Touch DONE

  1. Set your frequency range

  2. Open calibration menu

    Go to CAL > CALIBRATE

  3. Perform SHORT calibration

    • Connect SHORT standard to Port 1
    • Touch SHORT
    • Wait for completion
  4. Complete calibration

    Touch DONE

For S21 measurements, you can calibrate with just THRU.

  1. Set your frequency range

    Go to STIMULUS and configure START and STOP frequencies.

  2. Connect cables in THRU configuration

    Connect your Port 1 cable directly to your Port 2 cable (or use a THRU standard).

  3. Open calibration menu

    Go to CAL > CALIBRATE

  4. Perform THRU calibration

    • Touch THRU
    • Wait for sweep to complete
  5. Complete calibration

    Touch DONE

The NanoVNA automatically detects Enhanced Response mode based on which calibration standards were performed.

You can also explicitly enable it:

  1. Go to calibration menu

    Touch CAL

  2. Select Enhanced Response

    Touch ENHANCED RESPONSE

    This toggles Enhanced Response mode on or off.

  3. Check the status indicator

    When Enhanced Response is active, the calibration indicator may show differently than full calibration.

Missing StandardImpact
No LOADDirectivity error not corrected
No SHORT (when using OPEN)Limited accuracy for low-impedance loads
No OPEN (when using SHORT)Limited accuracy for high-impedance loads

Practical impact:

  • Works well for measuring resonant antennas near 50 ohms
  • Less accurate for very low or very high impedances
  • Phase accuracy is reduced
Missing StandardImpact
No ISOLNPort crosstalk not corrected
No Port 1 calibrationSource match errors not corrected

Practical impact:

  • Good for measuring insertion loss of attenuators and filters
  • Less accurate for devices with poor return loss
  • Isolation measurements may show floor at -40 to -50 dB

Quick Reference: Minimum Calibration Requirements

Section titled “Quick Reference: Minimum Calibration Requirements”

Minimum: OPEN only or SHORT only

Better: OPEN + SHORT

Best: OPEN + SHORT + LOAD (full S11 calibration)

Use for: Antenna SWR, return loss measurements

When you need a fast antenna check without a full calibration kit:

  1. Set frequency range for your band

    Example: 144 MHz to 148 MHz for 2m band

  2. Calibrate with OPEN

    Remove the antenna connector and perform OPEN calibration.

  3. Connect your antenna

  4. Read SWR

    The SWR reading will be approximate but useful for tuning.

  5. For better accuracy

    If you have a SHORT standard (even a piece of wire soldered across an SMA connector), perform SHORT calibration as well.

MethodStandards RequiredS11 AccuracyS21 AccuracySpeed
No calibrationNonePoorPoorInstant
OPEN only1FairN/AFast
SHORT only1FairN/AFast
THRU only1N/AFairFast
OPEN + SHORT2GoodN/AFast
THRU + ISOLN2N/AGoodFast
Full SOLT4-5ExcellentExcellentModerate
  1. Check current calibration status

    The status bar shows which calibration is active.

  2. Recall a different calibration

    Go to CAL > menu to access saved calibrations.

  3. Reset and recalibrate

    Go to CAL > RESET to clear current calibration, then perform new calibration.